Certifications

Test Aticle: Media 2; TR853102
Purchase Order: 220362
Study Number: 1183693-S01
Study Received Date: 16 May 2019
Testing Facilty: Nelson Laboratories, LLC, 6280 S. Redwood Rd, Salt Lake City, UT 84123 US A
Test Procedure(s): Standard Test Protocol (STP) Number. STP0004 Rev 16
Deviation(s): None

Summary: The BFE test is performed to determine the filtration efficiency of test articles by comparing
the bacterial control counts upstream of the test article to the bacterial counts downstream. A suspension
of Staphylococcus aureus was aerosolized using a nebulizer and delivered to the test article at a constant
flow rate and fixed ai pressure. The challenge delivery was maintained at 1.7 - 2.7 x 10^3 colony forming
units (CFU) with a mean particle size (MPS) of 3.0 +- 0.3 μm. The aerosols were drawn through a six-
stage, viable particie, Andersen sampler for collection. This test method complies with ASTM F2101-14,
Annex B, and AS4381:2015.

The Deita P test is performed to determine the breathability of test articles by measuring the differential
air pressure on either side of the test article using a manometer, at a constant flow rate  The Delta P test
was designed to comply with MIL-M-36954C, Section 4.4.1.2, and complies with AS4381:2015,

All test method acceptance criteria were met. Testing was performed in compliance with US FDA good
manufacturing practice (GMP) regulations 21 CFR Parts 210, 211 and 820.

Test Side:  Side with Sticker
BFE Test Area:  ~40 cm^2a
BFE Flow Rate:  28.3 Liters per minute (L/min)
Delta P Flow Rate: 8 L/min
Conditioning Parameters: 85 +- 5% relative humidity (RH) and 21 +- 5°C for a minimum of 4 hours
Positive Control Average: 1.8 x 10^3 CFU
Negative Monitor Count: <1 CFU
MPS: 2.9μm
* A small portion of the sticker was included in the test area.

Study Director: Signature of Janelle Bentz
Study Completion Date: 28 May 2019
Barcode : 1193693-S01

801-290-7500 | nelsonlabs.com | sales@nelsonlabs.com bsm FRT0004-0001 Rev 20
Page 1 of 2
These results apply to the samples as received and relate only to the test article listed in this report. Reports may not be reproduced except in their entiery. Subject to NL terms and conditions at www.nelsonlabs.com Results:
Test Article Number: 1, Percent BFE(%): >99.9^a, Delta P (mm H2O/cm^2): 1.8, Delta P (Pa/cm^2): 18.0
Test Article Number: 2, Percent BFE(%): >99.9, Delta P (mm H2O/cm^2): 1.8, Delta P (Pa/cm^2): 18.0
Test Article Number: 3, Percent BFE(%): >99.9^a, Delta P (mm H2O/cm^2): 1.8, Delta P (Pa/cm^2): 17.2
Test Article Number: 4, Percent BFE(%): >99.9^a, Delta P (mm H2O/cm^2): 1.7, Delta P (Pa/cm^2): 17.0
Test Article Number: 5, Percent BFE(%): >99.9^a, Delta P (mm H2O/cm^2): 1.9, Delta P (Pa/cm^2): 18.6
* There were no detected colonies on any of the Andersen sampler plates for this test article.
The filtration efficiency percentages were calculated using the following equation:
c-1 C = Positive control average
%BFE = ((C-T)/C)x100 
C = Positive Control Average
T = Plate count total recovered downstream of the test article
Note: The plate count total is available upon request.

801-290-7500 | nelsonlabs.com | sales@nelsonlabs.com bsm FRT0004-0001 Rev 20
Page 1 of 2